Products

DBB-EXA ES with D-FAS
拡大する

DBB-EXA ES with D-FAS

Smart HD Built on Proven Innovation

DBB-EXA ES with D-FAS brings the trusted performance of our flagship HDF model, DBB-EXA, to the HD treatment for everyone. DBB-EXA ES delivers the essential functions that matter—advanced workflow, cost-effective operation, and intelligent monitoring.

🔹 The D-FAS reduces the physical and cognitive burden of routine dialysis work, enabling safer, more efficient care with fewer manual steps.
🔹 DBB-EXA series achieves greener dialysis with significant cost savings through efficient use of consumables—without compromising treatment quality or operational reliability.
🔹 The advanced monitoring capabilities support personalized and responsive patient care by providing real-time patient insights.

Contact us today to learn how DBB-EXA ES with D-FAS can support your facility’s growth.

D-FAS

Revolutionize Dialysis Workflow with D-FAS

 

D-FAS and patient cards streamline routine dialysis work such as priming, blood filling, emergency bolus, and rise back procedures. It significantly reduces staff workload and enhances operational efficiency by equalizing workload variation of dialysis facilities.

 

Key Benefits:

Reduced Staff Burden – Free up valuable time for healthcare professionals to focus on patient care.

Minimized Risk of Errors

– Ensures consistent, reliable performance.

Easier Staff Training – Simplified workflows reduce onboarding, training time, and cost.

Efficient use of consumables

Dialysis fluid priming for a Greener, Cost-Effective Dialysis

 

DBB-EXA series enables dialysis fluid priming using standard bloodline sets, eliminating the saline consumption and need for an additional infusion circuit. This innovative approach contributes to lower environmental impact and cost savings.

Key Benefits:

Lower Environmental Impact – Supports sustainable dialysis practices.

Cost Savings – Less consumable need helps facilities save costs while keeping the quality of treatment.

Reducing inventory management need – Less consumable need helps facilities reduce inventory management needs, including storage space.

 

Monitoring

Comprehensive Monitoring Availability for Optimal Dialysis Care

 

DBB-EXA series can be equipped with a full suite of monitoring technologies, including inflationary NIBP, Dialysis Dose Monitor (DDM), Blood Volume Monitor (Heamo Master), and Vascular Access Recirculation Rate Measurement (VARRM). These features provide real-time insights into patient and treatment status, supporting physicians to deliver precise, personalized care.

 

Key Benefits:

Ensure treatment adequacy – DDM with real-time monitoring and VARRM with screening capability ensure treatment adequacy.

Support blood pressure management – Haemo Master provides real-time monitoring without extra consumables.

Reducing patients stress due to the routine BP measurement – iNIBP achieve BP measurement with lower cuff pressure and shorter measurement time.

Disclaimer:

  • Product appearance and specifications are subject to change without prior notice.
  • Photographs of the product may appear differently due to variations in printing or display conditions.
  • Specifications may vary depending on local regulations and registration status. Please contact your local distributor for details.
  • Always read the accompanying operation manual before use.

General data

Dimensions H150 cm x W43 cm x D50 cm
Base: W45 cm x D61 cm
Weight Approx. 80 kg(incl. all options)
Water supply Pressure: 1 to 7 bar
Drain Minimum drain capacity: 800 mL/min average
Height: 50 cm maximum
Temp.: 90 ℃ maximum
Concentrate supply Pressure: 0 to 0.5 bar
Power supply 220 to 240 VAC , 50/60 Hz
110 to 120 VAC , 50/60 Hz
Battery Ni-MH battery 24 V / 3200 mAh
External connection port External output(Staff call)
External input 1, External input 2
Network
BPM start switch
USB 2.0 or 3.0
Screen 15 inch LCD

Hydraulic circuit

Dialysis fluid flow rate Setting range:
Single ETRF 300 to 800 mL/min
Double ETRF 300 to 700 mL/min
Dialysis fluid temperature Setting range: 33.0 to 40.0 ℃
Dialysis fluid conductivity Bicarbonate dialysis
Bicarbonate conductivity setting range: 2.3 to 7.0 mS/cm
Total conductivity setting range: 12.7 to 15.2 mS/cm
Acetate dialysis
Total conductivity setting range: 12.7 to 15.2 mS/cm
Blood leak detector Method: Optical
Sensitivity: 0.3 mL Blood / 1 L Dialysis fluid
(Blood: Hematocrit 32 ± 2 %; Dialysis fluid temperature: 37 ℃)
Ultrafiltration UF rate: 0.00: 0.10 to 4.00 L/h
UF accuracy:
±30 mL/h(at dialysis fluid flow rate 300 to 500 mL/min)
±0.1 % of the dialysis fluid flow rate
(at dialysis fluid flow rate 501 to 800 mL/min)
Dialysis Dose Monitor Measurement principle: Absorptiometry
Applicable Treatment mode: HD
Applicable Kt/V range: 0 to 3.0
Applicable Kt range: 0 to 300.0
Applicable K range: 0 to 999.9
Kt/V monitoring accuracy:
±0.1 (Kt/V 0 to 1)
±10 % (Kt/V 1 to 3)
eKt/V: 0 to 30
Applicable URR range: 0 % to 100 %
URR monitoring accuracy: ±5 %
Endotoxin retentive
filter (ETRF)
EF-02/D series

Accessories

Patient card MIFARE Classic 4K
Capacity: 4096 byte
Arterial pressure
monitoring
Measurement range: ー300 to +500 mmHg
Measurement accuracy: ±10 mmHg
Venous pressure
monitoring
Measurement range: ー300 to +500 mmHg
Measurement accuracy: ±10 mmH
Dialyzer inlet blood
pressure monitoring
Measurement range: ー300 to +735 mmHg
Measurement accuracy: ±10 mmHg
Single needle pressure Measurement range: ー200 to +600 mmHg
Measurement accuracy: ±10 mmHg
Air detector Method: ultrasonic waves
Sensitivity: 0.02 mL(normal air bubbles)
(at Blood flow rate: 250 mL/min)
0.0003 mL(microbubbles: blood/air mixture)
(at Blood flow rate: 250 mL/min)
Arterial blood pump Setting range: 40 to 600 mL/min
Flow rate accuracy:
Set value ±10 %
(inlet Pressure -150 mmHg <= P <= +150 mmHg)
Set value -20 to 0 %
(inlet Pressure -200 mmHg <= P < -150 mmHg)
Heparin pump Setting range: 0.0 to 9.9 mL/h
Output rate accuracy:
Set value ±10 %
Syringe type: 30 mL or 20 mL, 20 mL or 10 mL
Bolus volume: 0.0 to 9.9 mL
Blood Pressure
Monitor (BPM)
For standard BPM:
Pressure display range: 10 to 300 mmHg
Pressure display accuracy: less than ±3 mmHg
Measurement range:
Adult systolic blood pressure(SYS): 60 to 250 mmHg
Mean arterial pressure(MAP): 45 to 235 mmHg
Diastolic blood pressure(DIA): 40 to 200 mmHg
Pulse rate: 40 to 200 beats per minute
For iNIBP:
Pressure display range: 0 to 300 mmHg
Pressure display accuracy: Less than ±3 mmHg
Measurement range:
Adult systolic blood pressure(SYS): 40 to 280 mmHg
Mean arterial pressure(MAP): 10 to 280 mmHg
Diastolic blood pressure(DIA): 10 to 235 mmHg
Pulse rate: 30 to 200 beats per minute
Blood Volume
Monitor (BVM)
Measurement principle: Near-infrared reflection method
Applicable blood flow rate range: 40 to 600 mL/min
Applicable hematocrit range: 15 to 50 %
dBV monitoring accuracy: ±2.3 dBV%(Double needle)

Treatment options

Single needle
treatment
Single needle single pump treatment
SN control pressure:
Upper limit: +200 mmHg
Lower limit: +50 mmHg
UF profile 9 programmable profiles available
Conductivity profile 9 programmable profiles available

Cleaning program

Disinfection
and decalcification
50 % Citric acid
Peracetic acid (DIALOX)
Disinfection
and degreasing
Sodium hypochlorite solution
(Maximum 10 % chlorite concentration in disinfectant)
Decalcification Acetic acid
(Maximum 30 % acetic acid concentration in decalcification solution)

Disclaimer:
* Product appearance and specifications are subject to change without prior notice.
* Photographs of the product may appear differently due to variations in printing or display conditions.
* Specifications may vary depending on local regulations and registration status. Please contact your local distributor for details.
* Always read the accompanying operation manual before use.

pagetop
The Medical device and therapy related information on this website is aimed exclusively at healthcare professionals.
By the clicking on the button “Yes” you assure that you have taken note of this information and you are a healthcare professional.
Are you a healthcare professional ?